Contact Us

AC32M

Chemiluminescence Single Chamber Nitrogen Oxides (NO, NOX, NO2) Analyzer


  • Low level monitoring of NO-NO2-NOXfrom 0.4 ppb to 20 ppm
  • Ultra compact and light – rack 3U
  • SMD enhanced electronics
  • High performance metrology
  • Enhanced data logging functions

  • APPLICATIONS
  • EXCLUSIVE FEATURES
  • SPECIFICATIONS
  • CERTIFICATIONS

 

  • Ambient air monitoring
  • Indoor air monitoring
  • Mobile laboratory
  • NO2 measurement in medical gases
  • Continuous emissions monitoring (CEM) by dilution extraction

 

  • Graphic Liquid Crystal Display (LCD)
  • Interactive menu driven software with enhanced speed display
  • Real-time synoptic flow diagram display
  • User programmable ranges and average times
  • Auto-ranging
  • Temperature and pressure compensation
  • Automatic response time
  • Real time calibration graph
  • Built-in storage of one month average data (up to 6 months with the optional memory extension)
  • Built-in double serial interface (RS232/RS422) for remote control and maintenance

 

 

 

  • Ranges: 0-0.05 / 0.1/ 0.2 / 0.5 /1 / 2 / 5 / 10 /20 ppm or user selectable ranges
  • Lower detectable limit (2s): 0.4 ppb
  • Response time: automatic and programmable (minimum 30 sec)
  • Zero drift: less than 0.5 ppb / 24 h less than 1 ppb / 7 days
  • Span drift: less than 0.5 % / 24 h less than 1 % / 7 days
  • Linearity: +/- 1 % of F.S.

 

 

Complies with ISO 7996 and EN 14211 :2005 standards

  • United States :US EPA, automatic reference method RFNA-0202-146
  • France : INERIS, DRC-01-23445-AIRE N°245E-DGu/DR
  • France : ACIME, NF 379-02/01/04-01 (NF IE)
  • Germany : TÜV-Report: 936/21205818/C. 2007-04-20, Nr 75,S.4139-4140